APA
Faus Matoses, Vicente & Faus Llácer, Vicente & Aldeguer Muñoz, Álvaro & Alonso Pérez-Barquero, Jorge & Faus Matoses, Ignacio & Ruiz Sánchez, Celia & Zubizarreta Macho, Álvaro .A Novel Digital Technique to Analyze the Wear of CM-Wire NiTi Alloy Endodontic Reciprocating Files: An In Vitro Study.
ISO 690
Faus Matoses, Vicente & Faus Llácer, Vicente & Aldeguer Muñoz, Álvaro & Alonso Pérez-Barquero, Jorge & Faus Matoses, Ignacio & Ruiz Sánchez, Celia & Zubizarreta Macho, Álvaro. A Novel Digital Technique to Analyze the Wear of CM-Wire NiTi Alloy Endodontic Reciprocating Files: An In Vitro Study.
https://hdl.handle.net/20.500.12080/29540
dc.contributor.author |
Faus Matoses, Vicente |
|
dc.contributor.author |
Faus Llácer, Vicente |
|
dc.contributor.author |
Aldeguer Muñoz, Álvaro |
|
dc.contributor.author |
Alonso Pérez-Barquero, Jorge |
|
dc.contributor.author |
Faus Matoses, Ignacio |
|
dc.contributor.author |
Ruiz Sánchez, Celia |
|
dc.contributor.author |
Zubizarreta Macho, Álvaro
|
|
dc.date.accessioned |
2022-04-18T14:52:53Z |
|
dc.date.available |
2022-04-18T14:52:53Z |
|
dc.date.created |
2022-03-22 |
|
dc.identifier.uri |
https://hdl.handle.net/20.500.12080/29540 |
|
dc.description.abstract |
Abstract: The present study seeks to describe a novel digital measurement technique for analyzing
the wear volume of controlled memory (CM)-wire NiTi alloy endodontic reciprocating files after
clinical use. Material and Methods: Ten CM-wire NiTi endodontic reciprocating files were randomly
used in ten first upper molar teeth within four root canals. The CM-wire NiTi alloy endodontic
reciprocating files were subjected to preoperative and postoperative micro-computed tomography
(micro-CT) scans to obtain accurate digital imaging and communication on medicine (DICOM) digital
files, which were segmented using intensity-based thresholding and an exclusive OR (XOR) logical
operation (Boolean algebra logical operator) to obtain a mask of the location to localize and quantify
the wear volume of the CM-wire NiTi alloy endodontic reciprocating files. Gage repeatability and
reproducibility statistical analysis was applied to assess the reproducibility and repeatability of this
measurement technique. Results: The analysis showed a repeatability and reproducibility of 0.00%
for the digital measurement technique used to analyze the wear volume of CM-wire NiTi alloy
endodontic reciprocating files after clinical use. Wear was mostly identified at the tip and cutting
edges of the CM-wire NiTi alloy endodontic reciprocating files. Conclusions: This novel digital
measurement technique is a repeatable, reproducible, and accurate method of quantifying the wear
volume of CM-wire NiTi alloy endodontic reciprocating files after clinical use.
Keywords: endodontics; cyclic fatigue; wear; NiTi; reciprocating movement; micro-computed tomography |
es_ES |
dc.format |
application/pdf |
es_ES |
dc.language |
eng |
es_ES |
dc.rights |
CC-BY |
es_ES |
dc.rights.uri |
http://creativecommons.org/licenses/by/4.0/deed.es |
es_ES |
dc.title |
A Novel Digital Technique to Analyze the Wear of CM-Wire NiTi Alloy Endodontic Reciprocating Files: An In Vitro Study |
es_ES |
dc.type |
info:eu-repo/semantics/article |
es_ES |
dc.rights.accessrights |
info:eu-repo/semantics/openAccess |
es_ES |
dc.identifier.location |
N/A |
es_ES |