Faus Llácer, Vicente & Hamoud Kharrat, Nirmine & Marhuenda Ramos, María Teresa & Faus Matoses, Ignacio & Zubizarreta Macho, Álvaro & Ruiz Sánchez, Celia & Faus Matoses, Vicente .Influence of the Geometrical Cross-Section Design on the Dynamic Cyclic Fatigue Resistance of NiTi Endodontic Rotary Files¿An In Vitro Study.
ISO 690
Faus Llácer, Vicente & Hamoud Kharrat, Nirmine & Marhuenda Ramos, María Teresa & Faus Matoses, Ignacio & Zubizarreta Macho, Álvaro & Ruiz Sánchez, Celia & Faus Matoses, Vicente. Influence of the Geometrical Cross-Section Design on the Dynamic Cyclic Fatigue Resistance of NiTi Endodontic Rotary Files¿An In Vitro Study.
https://hdl.handle.net/20.500.12080/25980
Resumen:
The aim of this study was to analyze and compare the influence of the geometrical crosssection design on the dynamic cyclic fatigue resistance of NiTi endodontic rotary files. Materials and
Methods: Forty sterile endodontic rotary files were selected and distributed into the following study
groups: A: 25.06 double S-shaped cross-section NiTi alloy endodontic rotary files (Mtwo) (n = 10);
B: 20.04 rectangular cross-section NiTi alloy endodontic rotary files (T Pro E1) (n = 10); C: 25.04
convex triangular cross-section NiTi alloy endodontic rotary files (T Pro E2) (n = 10); and D: 25.06
triangular cross-section NiTi alloy endodontic rotary files (T Pro E4) (n = 10). A cyclic fatigue device
was used to conduct the static cyclic fatigue tests with stainless steel artificial root canal systems
with 200 µm and 250 µm apical diameter, 60¿
curvature angle, 3 mm radius of curvature, 20 mm
length, and 4% and 8% taper. The results were analyzed using the ANOVA test and Weibull statistical
analysis. Results: All the pairwise comparisons presented statistically significant differences between
the time to failure and number of cycles to failure for the cross-section design study groups (p < 0.001).
Conclusions: the double S-shaped cross-section of Mtwo NiTi endodontic files shows higher cyclic
fatigue resistance than the rectangular cross-section of T Pro E1 NiTi endodontic files, the convex
triangular cross-section of T Pro E2 NiTi endodontic files, and the triangular cross-section of T Pro E4
NiTi endodontic files.