Zubizarreta Macho, Álvaro & de Pedro Muñoz, Ana & Riad Deglow, Elena & Agustín Panadero, Rubén & Mena Álvarez, Jesús .Accuracy of Computer-Aided Dynamic Navigation Compared to Computer-Aided Static Procedure for Endodontic Access Cavities: An In Vitro Study.
ISO 690
Zubizarreta Macho, Álvaro & de Pedro Muñoz, Ana & Riad Deglow, Elena & Agustín Panadero, Rubén & Mena Álvarez, Jesús. Accuracy of Computer-Aided Dynamic Navigation Compared to Computer-Aided Static Procedure for Endodontic Access Cavities: An In Vitro Study.
https://hdl.handle.net/20.500.12080/24010
dc.contributor.author |
Zubizarreta Macho, Álvaro
|
|
dc.contributor.author |
de Pedro Muñoz, Ana
|
|
dc.contributor.author |
Riad Deglow, Elena
|
|
dc.contributor.author |
Agustín Panadero, Rubén |
|
dc.contributor.author |
Mena Álvarez, Jesús
|
|
dc.date.accessioned |
2021-02-09T16:46:24Z |
|
dc.date.available |
2021-02-09T16:46:24Z |
|
dc.date.created |
2020-01 |
|
dc.identifier.uri |
https://hdl.handle.net/20.500.12080/24010 |
|
dc.description.abstract |
Purpose: To analyze the accuracy of two computer-aided navigation techniques to guide
the performance of endodontic access cavities compared with the conventional access procedure.
Materials and Methods: A total of 30 single-rooted anterior teeth were selected, which were randomly
distributed into three study groups: Group A¿guided performance of endodontic access cavities
through computer-aided static navigation system (n = 10) (SN); Group B¿guided performance of
endodontic access cavities through computer-aided dynamic navigation system (n = 10) (DN); and
Group C¿manual (freehand) performance of endodontic access cavities (n = 10) (MN). The endodontic
access cavities of the SN group were performed with a stereolithography template designed on
3D implant planning software, based on preoperative cone-beam computed tomography (CBCT)
and a 3D extraoral surface scan, and endodontic access cavities of the DN group were planned and
performed by the dynamic navigation system. After endodontic access cavities were performed,
a second CBCT was done, and the degree of accuracy between the planned and performed endodontic
access cavities was analyzed using therapeutic planning software and Student¿s t-test. Results: Paired
t-test revealed no statistically significant di erences between SN and DN at the coronal (p = 0.6542),
apical (p = 0.9144), or angular (p = 0.0724) level; however, statistically significant di erences were
observed between the two computer-aided navigation techniques and the MN group at the coronal
(p < 0.0001), apical (p < 0.0001), and angular (p < 0.0001) level. Conclusion: Both computer-aided static
and dynamic navigation procedures allowed accurate performance of endodontic access cavities. |
es_ES |
dc.format |
application/pdf |
es_ES |
dc.language |
eng |
es_ES |
dc.rights |
CC-BY |
es_ES |
dc.rights.uri |
N/A |
es_ES |
dc.title |
Accuracy of Computer-Aided Dynamic Navigation Compared to Computer-Aided Static Procedure for Endodontic Access Cavities: An In Vitro Study |
es_ES |
dc.type |
info:eu-repo/semantics/article |
es_ES |
dc.rights.accessrights |
info:eu-repo/semantics/openAccess |
es_ES |
dc.identifier.location |
N/A |
es_ES |
Fichier(s) constituant ce document
Ce document figure dans la(les) collection(s) suivante(s)
Afficher la notice abrégée